TPS QA Periodicities

TABLE 1

TPS QA TASKS AND PERIODICITY

   

TASK/CHECK

PROCEDURE

PERIODICITY

REFERENCE

Non-Dosimetric Evaluation – Digitizer Accuracy

1.1.2.2

 

Semi-Annually

 

TG- 53, IAEA TR 430, Note 1

Non –Dosimetric Evaluation – Image Acquisition, Use, and Display

1.1.2.3

1.1.2.3A

Semi-Annually

Quarterly

TG- 53, IAEA TR 430, Note 1,

AAPM Report 39

Ultrasound TG-1

Non-Dosimetric Evaluation – Hardcopy Output Accuracy

1.1.2.5

Semi-Annually

TG- 53, IAEA TR 430, Note 1

Dosimetric Evaluation – Monitor Unit Check – Open and Wedge Fields

1.1.2.4

Semi-Annually

TG- 53, IAEA TR 430, Note 1

Dosimetric Evaluation – Isodose Checks

1.1.2.4

Semi-Annually

TG- 53, IAEA TR 430, Note 1, Note 2

Dosimetric Evaluation – Clinical Isodose/Monitor Unit Check

1.1.2.4

Semi-Annually

TG- 53, IAEA TR 430, Note 1, Note 2

Dosimetric Evaluation –

Electron Monitor Unit and PDD Check

1.1.2.6

Semi-Annually

TG-53, IAEA TR 430

Dosimetric Evaluation –

Operating Consistency of IMRT Dose Optimization Software

1.1.2.7

Semi- Annually

TG-53, IAEA TR 430

 

ASTRO and ACR Practice Guidelines for IMRT

Dosimetric Evaluation- HDR Treatment Planning QA

1.1.2A.1

Semi-Annually

TG-53, IAEA TR 430

Dosimetric Evaluation - Prostate Seed Treatment Planning QA

1.1.2B.1

Semi-Annually

TG-53, IAEA TR 430

 

 

TABLE 1- NOTES

 

Note 1: TG-40 Section III and Table V identifies suggested periodicity and test criterion for this activity. The semi-annual activity will be performed by the dosimetrist per Procedures 1.1.2.2, 1.1.2.3, 1.1.2.4, 1.1.2.5, and 1.1.2.6. The physicist will review and approve the results.

 

Note 2:  QA for Treatment Planning Dose Delivery by 3DRTP and IMRT; Boyer, et al.; General Practice of Radiation Oncology Physics in the 21st Century: Med Phys Publishing 2000.  This reference has suggested criterion for this activity.